Using a computational method based on static equilibrium, it is possible to independently measure the rod's
orientation and tip position during the movement trial, using force platform derived data exclusively, and to
compare these estimates to rod orientation and tip position estimates derived exclusively from the motion capture
component. The degree of spatial and temporal agreement between these two sets of data provides an assessment of
the joint working status of the kinetic and kinematic systems. A small number of variable are calculated and presented
for inspection—these enable the user to detect the presence of errors from a large number of potential sources.
Coming Soon:
CalTester Version 4.0 will incoproate the functionality of FPLOC to locate force platform corners, and more
importantly will allow editing of the Force Platform parameters in the C3D file.
|